34
2. Press the
INSER
button to confirm.
3. If the piece entered is less than 100 times the resolution of the balance, an error message will
be displayed.
To exit without entering the weight, press
ESCAPE
(on the alphanumeric keyboard) or
ON/OFF
.
4.
If the weight is sufficient, “
0
” will be shown on the display; it is now possible to proceed with
the counting, loading the pieces on the plate.
5. To exit from the piece counting function, press the
ON/OFF
button.
10.1.2 Automatic updating of the average unit weight
After having carried out the sampling, the average unit weight can be updated in the following
way.
1. Instead of loading all of the pieces to be counted, load a number of pieces approximately
double that of those loaded on the plate and wait for the beep.
2. This procedure can now be repeated up to a maximum of 255 pieces or you can proceed with
the normal counting of the pieces.
This mechanism allows for a more accurate estimate of the average unit weight and a better
precision in the counting of the pieces.
NOTE: the automatic updating mechanism is not active if the sampling has been carried out through
insertion of the average unit weight.
10.2 Program for the determination of the density of a solid or a liquid
The density calculation program allows the determination of the density of a solid or liquid
through the use of the lower weighing hook or the hydrostatic kit (optional Code T027).
10.2.1 Solid density determination
1. Select the density program as described in paragraph 10.
The following screen will be shown on the display:
2. Then press the
button to confirm the selection.
Density