![Accuris Dx W3101A-220 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/accuris/dx-w3101a-220/dx-w3101a-220_operation-manuals_456685010.webp)
10
2. When the value of the calibration weight begins to flash, load the weight indicated by the
display on the plate.
3. The display will stop flashing, indicating the value of the calibration weight with the stability
indicator on.
Once the calibration has been carried out, the calibrated weight will be displayed with the
indication of the current unit of measurement.
4. Remove the calibration weight.
The balance is ready for weighing operations.
NOTE: an error message will be displayed if there is any interference during the calibration process. To
interrupt the calibration process, press the ON/OFF button while the calibration weight indication flashes.
It is also possible to calibrate the balance with a calibration weight greater than the pre-set
calibration weight:
1. Press and hold the
CAL
button with the weigh pan empty until the beeping stops, and then
release the button. The word
“-CAL-“
is be shown on the display, followed by the word
“LOAD”
, flashing.