61
12 Error codes
ERROR MESSAGE
ON DISPLAY
MEANING
POSSIBLE SOLUTIONS
ERR01
Weight not stable after operation
of tare
Protect the balance from air
flows or from vibrations
from the table.
ERR02
impossible to start the calibration
due to instability of the balance
Protect the balance from air
flows or from vibrations
from the table.
ERR03
calibration weight not correct or
balance unstable
Calibrate with correct
weight or protect the
balance from environmental
disturbances.
In models
with internal calibration
remove the screw in the left
lower part of the balance
(see part 5)
ERR04
weight of samples for the piece
counting function not adequate or
unstable
Select a larger number of
samples or protect the
balance from vibrations.
ERR05
impossible to print because of
unstable weight
Protect the balance from
environmental disturbance
ERR06
Weight is not stable in density
mode
Protect the balance from
environmental disturbance
ERR07
Weight is not stable in percent
weighing mode
Protect the balance from
environmental disturbance
ERR08
Anomaly with autocalibration
motor
Contact service staff
ERR09
Weight is not stable in formulation
mode
Protect the balance from
environmental disturbance
ERR10
Weight of component out of
tollerance in formulation mode
Reduce quantity
ERR F
Flash memory damaged
Ask assistance from
authorized service staff
“UNLOAD”
weight loaded on the pan or pan
not positioned properly
Remove the weight from
the pan or position properly
the pan and underpan.
“CAL But”:
the balance requires to be re-
calibrated
Unload weights, if any, on
the pan, and press the CAL
button
Over-load condition
Unload the weights loaded
on the pan
Under-load condition
Place properly pan and
underpan
Содержание Dx W3101A-220
Страница 60: ...60 TEST Weight 18 71 g Signature G L P parameters Weight data ...