11.4.3 On demand transmission with G.L.P.
In the on demand transmission with G.L.P., the data transmitted to computer are the same as
described as in the on demand transmission without G.L.P. mode but with the add of G.L.P.
parameters before of each transmission, as described below:
--------------------
03-04-11 12:14:03
Balance ID:
A99
User ID:
MR.ROSSI
Project ID:
TEST
--------------------
Weight: 18.71 g
Signature:
--------------------
11.5 Connecton of balance with serial printer
It is possible to connect the balance to a printing peripherical.
To print the weight, connect the connector of the balance to the serial printer as shown in the
scheme below:
Here you can find description of the several modes of printing that can be selected:
•
Print of weight data with a generic serial printer (from the menu, set the generic printing
mode as described in the paragraph 9.2 and manage the busy signal).
•
Print of weight data together with GLP indications with generic serial printer (from the
menu, set the generic printing-GLP mode as described in the paragraph 9.2 and manage
the busy signal)
•
Print of weight data with printer model TLP50 (from the menu, set the printer TLP mode
as described in the paragraph 9.2).
•
Print of weight data together with GLP indications with printer model TLP50 (from the
menu, set the printer TLP - GLP mode as described in the paragraph 9.2).
Note: In all different printing modes just described, if the weight is not stable during transmission
of data to printer, an acoustic signal is emitted and ERR05 is displayed and weight is not printed.
G.L.P. parameters
G.L.P. parameters
Weight data
2
4
5
Input data
Busy signal
GND
Balance
side
Printer side