5. Once the internal calibration has been stored, the balance will return to the normal weighing
condition.
6. Now re-enter the calibration menu as described in paragraph 9.1.12 and set the desired
internal, automatic, or external calibration modality.
9.13 Calibration data
This function allows you to display the data relative to the last calibration carried out.
▪
Date
▪
Calibration mode
▪
Correction
1. Select the calibration data parameter as described in paragraph 10.
The data relative to the last calibration carried out will be shown on the display:
2. Press the
button to print the calibration data.
3. Press the
ON/OFF
button to exit from the screen and return to the balance parameters menu.
It will now be possible to select another parameter or return to weighing mode by pressing the
ON/OFF
button
.
Calibration data
10-06-10
External calib. 2000.0g
Corr.: 1.2g
WARNING: this procedure must be carried out only using class E2 or equivalent
reference masses.