48
Press the green
DO
key to initiate the first measurement. The display should read
You may now reposition the sample for the second measurement in the average.
Press the green
DO
key to initiate the second measurement. The display should read
You may now reposition the sample for the third measurement in the average
Press the green
DO
key to initiate the third measurement. This will complete the
measurement, the measurement data will be averaged and, stored and the display will
show whatever measurement result is selected. Measurement error or warning messages
will occur in the same manner as in the
Ready-to-Measure
state. If you use the square
MODE
key to accept a measurement it will be included in the average. If you use the
DO
key or allow the message to timeout, the measurement will not be included in the
average.
The menu item
Adjust Avg Cnt
allows you to modify the number of individual
measurements in an average. This may be any number between -99 and +99 excluding
minus one, zero, and one. Negative numbers are used for automatic averaging. When the
number to average is negative the instrument will proceed from measurement to
measurement without waiting for the operator to initiate each measurement.
To select a new average count starting from the
Ready-to-Measure or Ready-to-
Average
state follow the path
READY-to-MEASURE or READY-to-AVERAGE
MODE to Pick
UP to Setup
UP to Service Menu
UP to Battery Check
UP to Average
MODE to Average On
UP to Adjust Avg Cnt
If you now press the square
MODE
key, and select
Adjust Avg Cnt
display will read
001 Rdy/Average
01 of 03
001 Rdy/Average
02 of 03