So long as there is a measurement stored in the instrument you should be able to use the
arrow keys to get to a display of the form
or
Color coordinates are displayed to two places past the decimal and the decimal is not
displayed. Thus for the color in the examples above the Lab is (91.10,1.57,26.28) and the
XYZ is (88.25,78.66,17.47).
Averaging multiple measurements
In some cases it is desirable to make multiple measurements of several points on a
sample and to average the results. Your instrument provides for this procedure.
Starting from the
Ready-to-Measure
state follow the path
READY-to-MEASURE
MODE to Pick
UP to Setup
UP to Service Menu
UP to Battery Check
UP to Average
MODE to Average On
MODE to Ready-to-Measure
The display should read
As in the
Ready-to-Measure
state, the number in the upper left hand corner will be the
next storage location. The second line gives the number of averages in each measurement
and the number completed. For this unit setup there will be three measurements in each
average and none (00) have been completed.
#011 L*=9110
a*=0157 b*=2628
#011 X=8825
Y=7866 Z=1747
001 Rdy/Average
00 of 03