Prepared By: Peter Brandt Version: 08 – 20080415
Page
26 of 26
40-6640 Service Manual
Hairspring wire is broken
Replace the hairspring wire
Connecting wires between
circuit boards #1 and #4 are broken
Repair connecting wires.
Core connection wire to laser is cut
Replace the connection wire
#1 Main Contol circuit board
does not work.
Replace circuit board.
5
No laser
Laser light source is broken
Replace the laser
Buzzer is broken
Replace buzzer
#4 Alarm Loop Circuit Board is loose
or not aligned properly
Adjust and tighten circuit board
Connecting wires between
circuit boards #1 and #4 are broken
Repair connecting wires.
6
No buzzer (when
instrument is tilted
beyond tolerance)
#1 Main Contol circuit board
does not work.
Replace circuit board.
Connecting wires between
circuit boards #1 and #6 are broken
Repair connecting wires.
#6 Alarm Switch circuit board
does not work.
Replace circuit board.
7
On-Off keypad for Alarm
Switch does not work
#1 Main Contol circuit board
does not work.
Replace circuit board.
See reasons listed in above (No. 7)
Repair accordingly
8
On-Off keypad indicator
light for Alarm Switch
does not work
LED of circuit board #2 is damaged
Replace LED