Display Shows Possible Cause(s)
Possible Solution(s)
Bolus Delivery
Unavailable
j
The meter cannot
communicate with the pump.
j
Retry or set to Manual Pump.
j
Ensure the meter and pump have
Bluetooth
wireless
technology turned on.
j
Ensure the meter and pump are within
communication range.
j
If the low battery icon is displayed on the meter, the
meter cannot communicate with the pump. Replace
the batteries.
j
The pump is currently in Stop
mode.
j
Take the pump out of Stop mode.
j
The pump is currently
delivering a bolus.
j
Allow the current bolus delivery to complete before
delivering the next bolus.
Bolus Too High
j
The bolus is above the
allowed amount. The bolus
was set to 50 Units.
j
Check the accuracy of all entries. If necessary,
contact your healthcare professional.
Bolus Too Low
j
The insulin pump cannot
deliver a 0.1 Unit of multiwave
bolus.
j
The pump will adjust the bolus to 0.2 Units.
51638_00048010155_C.indb 250
6/11/13 3:58 PM
Содержание Aviva
Страница 2: ...LAST UPDATE 2012 08 51638_00048010155_C indb 2 6 11 13 3 57 PM ...
Страница 12: ...6 Glossary 293 Index 301 51638_00048010155_C indb 6 6 11 13 3 57 PM ...
Страница 44: ...38 51638_00048010155_C indb 38 6 11 13 3 57 PM ...
Страница 142: ...136 51638_00048010155_C indb 136 6 11 13 3 58 PM ...
Страница 250: ...244 51638_00048010155_C indb 244 6 11 13 3 58 PM ...
Страница 270: ...264 51638_00048010155_C indb 264 6 11 13 3 58 PM ...
Страница 312: ...306 V vibrate settings 185 187 W warnings list of 216 221 warranty 275 277 51638_00048010155_C indb 306 6 11 13 3 58 PM ...
Страница 313: ...307 51638_00048010155_C indb 307 6 11 13 3 58 PM ...
Страница 314: ...308 51638_00048010155_C indb 308 6 11 13 3 58 PM ...
Страница 315: ...LAST UPDATE 2012 08 51638_00048010155_C indb 2 6 11 13 3 57 PM ...