Display Shows Possible Cause(s)
Possible Solution(s)
E-57 Electronic
Error
j
An electronic error has
occurred or, in rare cases, a
used test strip was removed
and reinserted.
j
Turn off the meter and remove the batteries. Wait at
least 30 seconds prior to reinserting the batteries.
Turn on the meter and perform a blood glucose or
control test. If the problem persists, contact
ACCU-CHEK Customer Care Service Center at
1-800-688-4578.
E-58 Temp. Error
j
The temperature is above or
below the proper range for
the meter.
j
Move the meter to an area within the proper
temperature range indicated for test strip use in the
test strip package insert. Wait 5 minutes before
turning on the meter. Repeat the test. Do not
artificially heat or cool the meter.
E-59 Battery
Empty
j
The batteries are extremely
low.
j
Insert new batteries. See Chapter 8, Maintenance.
E-60 Time/Date
Error
j
The time and date settings
may be incorrect.
j
You have changed the
batteries.
j
Ensure the time and date are correct and adjust, if
necessary. See Chapter 5, Changing Meter
Settings.
51638_00048010155_C.indb 257
6/11/13 3:58 PM
Содержание Aviva
Страница 2: ...LAST UPDATE 2012 08 51638_00048010155_C indb 2 6 11 13 3 57 PM ...
Страница 12: ...6 Glossary 293 Index 301 51638_00048010155_C indb 6 6 11 13 3 57 PM ...
Страница 44: ...38 51638_00048010155_C indb 38 6 11 13 3 57 PM ...
Страница 142: ...136 51638_00048010155_C indb 136 6 11 13 3 58 PM ...
Страница 250: ...244 51638_00048010155_C indb 244 6 11 13 3 58 PM ...
Страница 270: ...264 51638_00048010155_C indb 264 6 11 13 3 58 PM ...
Страница 312: ...306 V vibrate settings 185 187 W warnings list of 216 221 warranty 275 277 51638_00048010155_C indb 306 6 11 13 3 58 PM ...
Страница 313: ...307 51638_00048010155_C indb 307 6 11 13 3 58 PM ...
Страница 314: ...308 51638_00048010155_C indb 308 6 11 13 3 58 PM ...
Страница 315: ...LAST UPDATE 2012 08 51638_00048010155_C indb 2 6 11 13 3 57 PM ...