251
9
Display Shows Possible Cause(s)
Possible Solution(s)
Carbohydrate Ratio
j
Carbohydrate ratio is outside
of the acceptable meter
range.
j
Check your entries and contact your healthcare
professional to determine the appropriate settings.
Code Key Missing
j
The black code key is not
inserted.
j
Turn off the meter and insert the black code key.
If you need a black code key, please contact
ACCU-CHEK Customer Care Service Center at
1-800-688-4578.
Communication
Lost
j
Pump is out of range of the
meter.
j
Bolus delivery continuing, see pump screen.
j
Ensure the meter and pump have
Bluetooth
wireless
technology turned on.
j
Ensure the meter and pump are within
communication range.
j
If the low battery icon is displayed on the meter, the
meter cannot communicate with the pump. Replace
the batteries.
j
Use the pump to monitor or cancel a bolus which is
in the process of being delivered.
51638_00048010155_C.indb 251
6/11/13 3:58 PM
Содержание Aviva
Страница 2: ...LAST UPDATE 2012 08 51638_00048010155_C indb 2 6 11 13 3 57 PM ...
Страница 12: ...6 Glossary 293 Index 301 51638_00048010155_C indb 6 6 11 13 3 57 PM ...
Страница 44: ...38 51638_00048010155_C indb 38 6 11 13 3 57 PM ...
Страница 142: ...136 51638_00048010155_C indb 136 6 11 13 3 58 PM ...
Страница 250: ...244 51638_00048010155_C indb 244 6 11 13 3 58 PM ...
Страница 270: ...264 51638_00048010155_C indb 264 6 11 13 3 58 PM ...
Страница 312: ...306 V vibrate settings 185 187 W warnings list of 216 221 warranty 275 277 51638_00048010155_C indb 306 6 11 13 3 58 PM ...
Страница 313: ...307 51638_00048010155_C indb 307 6 11 13 3 58 PM ...
Страница 314: ...308 51638_00048010155_C indb 308 6 11 13 3 58 PM ...
Страница 315: ...LAST UPDATE 2012 08 51638_00048010155_C indb 2 6 11 13 3 57 PM ...