256
Display Shows Possible Cause(s)
Possible Solution(s)
E-54 Not Enough
Sample
j
Not enough blood or control
solution was drawn into the
test strip for measurement or
was applied after the test has
started.
j
Discard the test strip and repeat the test.
E-55 Code Key
Expired
j
The white code key is from an
expired lot of test strips.
j
This message may appear when using a white code
key in the meter. It means the test strips expire at
the end of the current month. At the end of the
month, discard the white code key and any
remaining test strips. Insert a black code key. Make
sure the time and date in the meter are correct. If
you need a black code key, please contact
ACCU-CHEK Customer Care Service Center at
1-800-688-4578.
E-56 Sample
Applied Early
j
Blood or control solution was
applied to the test strip before
the Apply Sample screen
appeared on the display.
j
Discard the test strip and repeat the test with a new
test strip.
51638_00048010155_C.indb 256
6/11/13 3:58 PM
Содержание Aviva
Страница 2: ...LAST UPDATE 2012 08 51638_00048010155_C indb 2 6 11 13 3 57 PM ...
Страница 12: ...6 Glossary 293 Index 301 51638_00048010155_C indb 6 6 11 13 3 57 PM ...
Страница 44: ...38 51638_00048010155_C indb 38 6 11 13 3 57 PM ...
Страница 142: ...136 51638_00048010155_C indb 136 6 11 13 3 58 PM ...
Страница 250: ...244 51638_00048010155_C indb 244 6 11 13 3 58 PM ...
Страница 270: ...264 51638_00048010155_C indb 264 6 11 13 3 58 PM ...
Страница 312: ...306 V vibrate settings 185 187 W warnings list of 216 221 warranty 275 277 51638_00048010155_C indb 306 6 11 13 3 58 PM ...
Страница 313: ...307 51638_00048010155_C indb 307 6 11 13 3 58 PM ...
Страница 314: ...308 51638_00048010155_C indb 308 6 11 13 3 58 PM ...
Страница 315: ...LAST UPDATE 2012 08 51638_00048010155_C indb 2 6 11 13 3 57 PM ...