40
Temperature Conditions for Direct Contact
Conditions for Safe Contact – Time
Accessible Part
Contact Time Limit
Touchscreen
Less than 10 minutes
Plastic enclosure of device
Less than 10 minutes
Button
Less than 10 minutes
Conditions for Safe Contact – Temperature
Accessible Part
Maximum Temperature while Device is in
use (ambient temperature 40°C or 104°F)
Touchscreen
46.1°C or 114.98°F
Plastic enclosure of device
47.0°C or 116.60°F
Button
42.3°C or 108.14°F
Touching the surface of the device with broken skin may aggravate a wound.
Infants or high-risk groups should not touch the surface of the device if there is a chance of burning the skin.