SENSORY NERVE BLOCK
Provides symptomatic relief of superficial and deep pain from localized dermal or segmental origin. Blocks pain
by causing a temporary nerve block through reactive depolarization (Conduction block) of the pain signal on its
way to the spinal input. Also known as Wedensky inhibition. Duration of relief is typically from 1 to 2 hours.
Faster onset of relief usually within 5 to 10 minutes. Target tissue is superficial and deep. Bipolar placement
over local nerve. Quadripolar placement over area of local pain, or at involved spine segment. The intensity
should be set to elicit a numb-gripping sensation just under muscle contraction.
User
Adjustable
Parameter
(Y/N)
Order in
Parameter
Set Loop
Parameter
Name
Adjustable
Range (unit)
Default (unit)
N
CARRIER FREQ
10.0 (kHz)
Y
1
TREATMENT TIME
0…99 (min)
15 (min)
Neuro Re-Ed
A.
UE TRIPHASIC PENS
This program is used to re-educate upper extremity agonist - antagonist muscle pairs where fine motor
control and positioning is required. It is called triphasic because the agonist fires first, followed by the
antagonist followed by a small burst of agonist muscle to slow down the antagonist and position the joint.
Patterned Electrical Neuromuscular Stimulation (PENS) is a precise reproduction of EMG firing patterns of muscle
agonist and antagonist pairs. The stimulation helps to re-establish proper firing sequence in the spinal and
supraspinal central pattern generators, re-educating neuromuscular function. PENS helps re-boot the sense position
system and neuromuscular junction to improve recruitment and muscle function. It should be used prior to Therex
to improve recruitment and motor performance. It’s also used to decrease spasticity and muscle tone.
User
Adjustable
Parameter
(Y/N)
Order in
Parameter
Set Loop
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y
2
PHASE DURATION
40…100 (μs)
70
(μs)
N
PULSE RATE
50 (Hz)
Y
1
TREATMENT TIME
0…99 (min)
15 (min)
N
CHANNEL MODE
Tri Phasic
B.
UE BIPHASIC PENS
This program is used to re-educate upper extremity agonist - antagonist muscle pairs to enhance reciprocal
and fast movement. It is called biphasic because the agonist fires first followed by the antagonist. Due to
reciprocating action of the muscle groups there is no third phase. Functional task simulation is reciprocal
movement such as typing, boxing, piano playing, swimming, upper extremity positioning speed walking,
rowing, etc. Patterned Electrical Neuromuscular Stimulation (PENS) is a precise reproduction of EMG firing
patterns of muscle agonist and antagonist pairs. The stimulation helps to re-establish proper firing sequence in the
spinal and supraspinal central pattern generators, re-educating neuromuscular function. PENS helps re-boot the
sense position system and neuromuscular junction to improve recruitment and muscle function. It should be used
prior to Therex to improve recruitment and motor performance. It’s also used to decrease spasticity and muscle tone.
User
Adjustable
Parameter
(Y/N)
Order in
Parameter
Set Loop
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y
2
PHASE DURATION
40…100 (μs)
70 (μs)
N
PULSE RATE
50 (Hz)
Y
1
TREATMENT TIME
0…99 (min)
15 (min)
Y
3
CYCLE TIME
2.5; 1.5; .75; .6; .5 (sec)
.75 (sec)
N
CHANNEL MODE
Bi Phasic