Figure 7
UNIT OUTPUT TEST
When testing a unit for output, the output tester should be used on a constant output protocol.
Power unit ON.
Insert output tester into channel A.
Select Pain / IFC Sensory.
Increase the output level to 20mA on channel A.
Observe the output tester LED.
–
If the output tester has a solid, brightly lit LED, then
there is sufficient output present on the selected
channel.
Change the output tester to channel B and follow the above
process.
Service Center
For repair or service of ACP Products and accessories, please call (800) 350-1100 and follow the prompts. Normal
hours of operation are 6:00am to 5:00pm Pacific Standard Time. After hours, please leave a message and a
technician will return your call during the next scheduled workday.
Patient-end of lead
wire, Red-coded
Patient-end of lead
wire, Black-coded
Device-end of
lead wire
LED lights green when lead
wire is functional, and lights
red when lead wire needs
to be replaced.