FAST CYCLE
Program simulates fast cycling, at 60 rpm, and can be used to retrain reciprocal movement of the L.E. with
or without resistance. Electrodes are applied to both quads while sitting with hips and knees flexed. Use a
lower extremity ergometer or pedal exerciser in combination with stimulation at mild to moderate
contraction intensity for closed chain functional training. Increase resistance as the patient progresses in the
rehab program.
User
Adjustable
Parameter
(Y/N)
Order in
Parameter
Set Loop
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y
2
PHASE DURATION
40…100 (μs)
70 (μs)
N
CYCLE TIME
1.0 (s)
N
PULSE RATE
50 (Hz)
Y
1
TREATMENT TIME
0…99 (min)
15 (min)
C.
SLOW WALK
Program simulates slow walking at a speed of 42 strides/minute. The program is used to re-educate gait
function. Electrodes are applied to both quads while sitting with hips and knees flexed. A light resistance
using weight cuffs or exercise band can be applied. The stimulation should cause a light to moderate
contraction. Transition to standing, with weight shifting activities then to walking in place in later rehab.
User
Adjustable
Parameter
(Y/N)
Order in
Parameter
Set Loop
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y
2
PHASE DURATION
40…100 (μs)
70 (μs)
N
CYCLE TIME
1.40 (s)
N
PULSE RATE
50 (Hz)
Y
1
TREATMENT TIME
0…99 (min)
15 (min)
D.
MEDIUM WALK
Program simulates medium-speed walking at a rate of 60 strides/minute. The program is used to re-educate
gait function. Electrodes are applied to both quads while sitting with hips and knees flexed. A light
resistance using weight cuffs or exercise band can be applied. The stimulation should cause a light to
moderate contraction. Transition to standing, with weight shifting activities then to walking in place in
later rehab.
User
Adjustable
Parameter
(Y/N)
Order in
Parameter
Set Loop
Parameter
Name
Adjustable
Range (unit)
Default (unit)
Y
2
PHASE DURATION
40…100 (μs)
70 (μs)
N
CYCLE TIME
1.0 (s)
N
PULSE RATE
50 (Hz)
Y
1
TREATMENT TIME
0…99 (min)