HAND CONTROL:
MFAC Hand Control:
Modified Sine wave into physiologic load 2 KHz, 2.5 KHz, 4 KHz , 5 KHz AND 10 KHz carrier
frequency
Burst rate adjustable from 0.1 to 999bps, 50% duty factor
Rate scan from 0-50% programmable with scan time, programmable from 0-20 seconds
LVPC Hand Control:
Asymmetrical biphasic square wave
Phase duration adjustable from 10 - 300 µs
Pulse rate adjustable from 0.1 – 250 pps
TIMER FUNCTIONS:
Treatment Timer:
Adjustable for 1-99 minutes in one minute increments. Turns output to zero and system off and sounds 10-
second buzzer to indicate completion of treatment.
Channel Timing:
In alternating mode channels follow each other sequentially. In simultaneous modes the output is
simultaneous. In delayed mode channel B is delayed.
TIMING SELECTIONS:
Ramp Time:
Adjustable from 0.1 to 9.9 seconds in 0.1-second increments. Ramps overlap in NMES Alt mode or with
channel delay.
ON Time:
Adjustable from 1 to 30 seconds in 1-second increments
OFF Time:
Adjustable from 0 (continuous) to 199 seconds in 1-second increments
PATIENT SAFETY SYSTEMS:
Activation:
Patient safety hand control shuts down output. Output modality may not be changed during operation. Output
levels are reset to zero at the start and completion of treatment.
MISC.:
Certificates and Approvals:
Devices are designed to meet or exceed all safety requirements of a medical device in its class per IEC
60601 and CSA C22.2 No. 601.1
CAUTION:
Federal law restricts this device to sale by or on the order of a physician (or other health
practitioner licensed by their State).
ACP reserves the right to change technical specifications and product availability without notice.