BiWaze Cough - User Reference Manual
24
4.3.2
Internal Battery
ABM offers an internal Lithium-ion battery pack for the device. Please contact your
sales support or health care provider for battery replacement.
Internal Battery can power the device with active therapies for up to 1 hour* on full
charge or equivalent to four 15 minutes therapies.
*
subjected to default settings, the actual run time can vary depending on age of the battery,
settings and actual active therapy time.
The user can call health care provider to replace the batteries or continue to use the
device on AC mains.
The internal batteries can charge simultaneously while the device is operating on the
AC mains power and switches to Battery power source when AC mains is
disconnected.
Disposal
: Do not dispose the batteries in landfill.
4.3.3
Device Power Source Indicators
Power source indicators are presented on the device and the display screen. These
indicators are described in detail below.
4.3.3.1
AC Power Indicators
When AC power is applied to the device and the display is off, the red AC LED
indicator on the Power On/Power Off button lights. When AC power is applied and
the display is on, the charging indicator on the battery symbol on top menu
ribbon appears. The Battery charging indicator turns off when the device is run on
battery power.