The therapy screen shows the text that trigger is on and allows 30 secs timeout in
pause phase to detect patient effort. If patient effort is not detected in that time
frame the treatment is paused.
Inspiratory Flow:
While the expiratory switching is desired to be fast and at high flow
to simulate cough, the inspiratory flow may be controlled to a comfortable level by
selecting the flow between low, medium and high.
The High Flow setting applies maximum flow to target the Inspiratory pressure as
fast as possible providing maximum peak volumetric flow based on set pause and
inspiratory pressure settings.
At Medium setting, the flow is controlled in such a way that the peak volumetric flow
observed during Inspiratory phase is within 70% to 40% of that observed when High
setting is applied.
At Low setting, the flow during Inspiratory phase is controlled in such a way that the
peak volumetric flow observed during Inspiratory phase is within 40% to 10% of that
observed when High setting is applied.