BiWaze Cough - User Reference Manual
59
Accuracies
stated
in
this
manual
are
based
on
specific
environmental
conditions.
For
stated
accuracy,
the
environmental
conditions
are:
Temperature:
20-30°
C;
Humidity:
50%
relative;
Altitude:
nominally
380
meters.
12.8
Control Accuracy
Parameter
Range
Accuracy
Pressure
-70
to
70
cmH
2
O
±
5
cmH
2
O
Inhale
Time
0-5
seconds
±
(10%
of
setting
+
0.1
second)
Exhale
Time
0-5
seconds
±
(10%
of
setting
+
0.1
second)
Pause
Time
0-5
seconds
±
(10%
of
setting
+
0.1
second)
Frequency
1-20
Hz
±
(10%
of
setting)
Amplitude
1-10
cmH
2
O
±
5
cmH
2
O
Device
performance
and
accuracy
is
specified
at
Temperature:
20-30°
C;
Humidity:
50%
relative;
Altitude:
nominally
380
meters
for
typical
patients.
12.9
Sound
The sound pressure of the device set at -40 cmH2O/+40 cmH2O in the Pause phase
is less than 65 dBA at 1 meter.
12.10
Disposal
Dispose of this device in accordance with local regulations. This device should be
disposed of separately, not as unsorted municipal waste. To dispose of your device,
you should use appropriate collection, reuse and recycling systems available in your
region. The use of these collection, reuse and recycling systems is designed to
reduce pressure on natural resources and prevent hazardous substances from
damaging the environment
12.11
Essential Performance
Note
: The Essential Performance of the BiWaze Cough Device is defined as follows:
Содержание BiWaze Cough