SOTOX A™ MOBILE ANALYSER | USER GUIDE
13 | EN
6. RUNNING A DRUG TEST
6.
Running a Drug Test
Prior to collecting and testing a sample taken from a donor, the analyser will prompt you to complete QC testing. QC stands for Quality Control. Two
reusable QC test cartridges are supplied with the SoToxa Mobile Analyser:
• Positive Quality Control (PQC).
• Negative Quality Control (NQC).
These two cartridges are tested in sequence to check that the analyser is interpreting test cartridge results correctly. If the QC tests pass, this means that the
analyser is working correctly.
6.1
Initiating a drug test
Step 1: Refer to section
4 Start-up
to switch on the analyser. Check that both QC cartridges are clean and undamaged. If the QC cartridges are marked, please
contact Product Support.
Step 2: When the analyser is ready to begin a test, the screen below will appear:
Step 3: Insert the QC cartridge, the cartridge will be analysed automatically. The order in which the QC cartridges are inserted is not important, as the analyser is able
to recognise the cartridge that has been inserted and will analyse it accordingly.
Step 4: Once analysis of the PQC cartridge is finished, the system will prompt you to remove the current cartridge and insert the NQC cartridge.
Step 5: Following analysis of the second NQC cartridge, you will be prompted to start the drug test process. Remove the NQC cartridge.
Step 2
Step 3
Step 3
Step 4
Step 5