1-4
Questions?
Refer to Page 10-9 For Technical Support Instructions.
Chapter 1: Overview
Precision PCx Plus Blood Glucose Test Strips use proprietary glucose-specific chemistry that includes the glucose
dehydrogenase enzyme, NAD cofactor and PQ mediator (GDH-NAD/PQ). This chemistry is used for testing glucose
because it is not affected by maltose, icodextrin, and other common substances at normal therapeutic levels. It also ensures
minimal measurement bias from oxygen, hematocrit and other physiological variables.
Precision PCx Plus Blood Glucose Test Strips also have exclusive blood application features to ensure reliable sampling.
First, the target area may be filled by applying a drop of sample to the top of the test strip or by dipping the end of the test
strip in the sample. Second, the fill trigger electrode ensures that the test will only start once sufficient blood has been
applied. Together, these features minimize the possibility of errant results due to sample application technique.
IMPORTANT
: Do not apply blood to the test strip when the test strip is out of the monitor. Do not use wet, bent,
scratched or damaged test strips.
Each test strip comes wrapped in a foil packet bearing a barcode label. The individual foil packet maintains the integrity of
each test strip by protecting it from exposure to air and moisture, and accidental contamination. The bar code label holds
information about the test strip including the lot number, calibration, expiration date, and expected control solution ranges.
One quick scan provides the monitor with all this information at the time of the test.
IMPORTANT
: Do not scan a packet’s bar code and use a test strip from another packet. This may cause
incorrect results to be generated. Use the test strip immediately after opening its foil packet.
Do not use the test strip if the foil packet has a puncture or tear in it.
The Precision Xceed Pro monitor is compatible with Precision PCx Plus Test Strips. Precision PCx Plus Test Strips are
protected by a gold foil packet and are identified on the barcode label. Inside each box of test strips is a package insert with
detailed instructions for use.
Precision PCx Plus Blood Glucose Test Strip
IMPORTANT
: The Precision Xceed Pro System will only work with Precision PCx Plus brand test strips. Use of
any other test strip may cause erroneous results.
Control Solutions
Control solutions are used to perform regular quality control checks on the monitor to ensure it is functioning correctly.
Control testing is further described in Chapter 4. Precision, Optium, or MediSense® Control Solutions may be used with the
Precision Xceed Pro System. These control solutions are available in one, two or three level configurations.
Calibration Verification Controls
RNA Medical® brand Glucose Calibration Verification Controls may also be used to confirm the calibration and analytical
measurement range of the monitor. This is called linearity testing and requires a kit with 5 levels of control solution.
Linearity testing is further described in Chapter 7.
Содержание Precision XceedPro
Страница 21: ...4 Control Test Operating Guidelines for Control Solution Testing Control Test Procedure Glucose ...
Страница 35: ...6 Proficiency Test ...
Страница 39: ...7 Linearity Test ...
Страница 55: ...9 Data Upload Docking Station Connecting Cables to the Docking Station Mounting Instructions ...
Страница 71: ...11 Maintenance Cleaning the Exterior Surface Replacing the Port Protector Battery Life Replacing the Batteries ...