Routine Operation
Shutdown
60
PLEX-ID SP User Manual 10/2009 N09455
Abbott Laboratories
To shut down the PLEX-ID SP, follow the steps below:
1.
Switch OFF the PLEX-ID SP by pressing the power switch
(Figure 2–4) at the bottom left of the back/side panel of the
instrument into the OFF position.
Warning
Discard the plastic tip comb onto the plate. Dispose of all tip
combs as biohazardous waste.
Remove any plates still in the instrument. Dispose of all microplates as
biohazardous waste.
▲
2.
Wipe the turntable surface and the adjacent instrument surface with
a soft cloth or tissue paper moistened with distilled water, a mild
detergent (SDS, sodium dodecyl sulfate) or soap solution.
3.
If you have spilt infectious agents on the turntable, disinfect with
70% ethanol or some other disinfectant (see “Decontamination
procedure” on page 64).
In case there is any abnormal situation during operation, such as fluids
spilling inside the instrument, follow the steps below:
1.
Switch OFF the instrument (Figure 2–4).
2.
Unplug the instrument immediately from the power supply
(Figure 2–4).
3.
Carry out appropriate corrective measures. However, do not
disassemble the instrument.
4.
If the corrective measures taken do not help, contact authorized
technical service or your local Abbott Molecular representative.
Shutdown
Emergency
situations
Содержание PLEX-ID SP
Страница 1: ...PLEX ID SP User Manual Rev 1 0 ...
Страница 2: ......
Страница 3: ...PLEX ID SP User Manual Rev 1 0 Cat no N09455 ...
Страница 6: ......
Страница 70: ...Maintenance Disposal of the instrument 68 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Страница 78: ...Frequently Asked Questions Q As 76 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Страница 82: ...Troubleshooting Guide Service request protocol 80 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Страница 86: ...Glossary 84 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Страница 89: ...Abbott Laboratories PLEX ID SP User Manual 10 2009 N09455 87 Notes ...
Страница 90: ...88 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories Notes ...
Страница 91: ...Abbott Laboratories PLEX ID SP User Manual 10 2009 N09455 89 Notes ...
Страница 92: ......
Страница 93: ...Abbott Molecular Inc 1300 E Touhy Ave Des Plaines IL 60018 USA www abbottmolecular com N09455 ...