Technical Specifications
In conformity with the requirements
Abbott Laboratories
PLEX-ID SP User Manual 10/2009 N09455
71
The safety specifications are also met under the following environmental
conditions in addition to or in excess of those stated in the operating conditions:
Altitude
Up to 2000 m
Tempe5
°
C to +40
°
C
Humidity
Maximum relative humidity 80% for temperatures up to 31
°
C
decreasing linearly to 50% relative humidity at 40
°
C
Mains supply
fluctuations
±
10% from nominal
Installation category
(overvoltage category)
II according to IEC 60664-1 (see
Note
1)
Pollution degree
2 according to IEC 60664-1 (see
Note
2)
Note
1) The
installation category
(overvoltage category) defines the level of
transient overvoltage which the instrument is designed to withstand safely.
It depends on the nature of the electricity supply and its overvoltage
protection means. For example, in CAT II which is the category used for
instruments in installations supplied from a supply comparable to public
mains, such as hospital and research laboratories and most industrial
laboratories, the expected transient overvoltage is 2500 V for a 230 V
supply and 1500 V for a 120 V supply.
2) The
pollution degree
describes the amount of conductive pollution
present in the operating environment. Pollution degree 2 assumes that
normally only nonconductive pollution, such as dust, occurs with the
exception of occasional conductivity caused by condensation.
▲
Содержание PLEX-ID SP
Страница 1: ...PLEX ID SP User Manual Rev 1 0 ...
Страница 2: ......
Страница 3: ...PLEX ID SP User Manual Rev 1 0 Cat no N09455 ...
Страница 6: ......
Страница 70: ...Maintenance Disposal of the instrument 68 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Страница 78: ...Frequently Asked Questions Q As 76 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Страница 82: ...Troubleshooting Guide Service request protocol 80 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Страница 86: ...Glossary 84 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Страница 89: ...Abbott Laboratories PLEX ID SP User Manual 10 2009 N09455 87 Notes ...
Страница 90: ...88 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories Notes ...
Страница 91: ...Abbott Laboratories PLEX ID SP User Manual 10 2009 N09455 89 Notes ...
Страница 92: ......
Страница 93: ...Abbott Molecular Inc 1300 E Touhy Ave Des Plaines IL 60018 USA www abbottmolecular com N09455 ...