3.1C Commissioning Page
Fig. 3.3C Identifying the ACJC Transistor
Fig. 3.4C Identification of Module Types
Select the Commissioning Pag
.
N o t e .
It is possible to access the Electrical Calibration
procedures further
down this page using Security Access at this point – see Section 4.
Software Identification
Shows the software type and issue number to be quoted on any
correspondence with the Company.
Automatic Cold Junction Compensation (ACJC) Constant
The cold junction reference temperature (in °C) is shown on the upper display.
If the value is suspect it may be checked as follows:
a) Measure the temperature at the ACJC transistor (TR1) with a
mercury-in-glass thermometer – see Fig. 3.3C.
b) Adjust the displayed value to that of the measured temperature.
N o t e. Adjusting the ACJC constant affects the calibration.
COMMISSIONING
Software Identification
ACJC const & ENT
➞
➞
Measure Temperature
▼
▲
or
Continued opposite.
Advance to next parameter.
Advance to next parameter.
3
1
TR1
C6
C35
R29
C8
IC1
IC2
ACJC Transistor