4.3C Setting Up the Analogue Retransmission
All analogue retransmission outputs are calibrated to the cell thermocouple input range. The input must be set up for a type K
thermocouple, with
°
C display units, to give the widest span (and hence the maximum accuracy), as follows:
SECURE PARAMETE
Cell Temperature, Display Units
Select °C display units.
Store.
Advance to next parameter.
Select the Display Temperature Page.
Temp
DISPLAY TEMPERATURE
Cell temp in °C or
➞
➞
°F or
Enter
Select the Commissioning Pag
.
Cell Thermocouple
Select cell thermocouple.
Store.
COMMISSIONING
Cell T/C? yes & ENT
➞
no & ENT
➞
Enter
Advance to cell thermocouple parameter.
3 Times
▼
▲
or
Enter
Number of Retransmission Outputs
Identify the number of retransmission modules fitted – see Fig. 3.4C on page
4.
Make a note of the retransmission range(s) and then reset it for 4 to 20mA
operation – see Section 6.2 in ZMT/0011.
Set the number of retransmission outputs being used (2 maximum on basic
analysers).
Store.
No of An Tx & ENT
➞
➞
▲
Operate for approximately three seconds to gain access to the SECURE
P A R A M E T E R S
within the Temperature Page.
➞
°C or
▲
Advance to SECURE
PARAMETERS
.
4 Times
➞
yes
▲
& ENT
Continued opposite.