8
Basic Analysers
Enter
Cell Thermocouple
Select whether or not a cell thermocouple is to be used:
If cell thermocouple is used (general applications).
or
If cell thermocouple is not used (special applications only).
N o t e. A preset cell temperature (>600°C) must be used if 'no' is selected.
Store.
Advance to Carbon Monoxide Signal
(basic analysers).
or
Advance to next parameter (advance analysers).
Flue Thermocouple
Select whether or not a flue thermocouple is to be used:
If flue thermocouple is used.
or
If flue thermocouple is never to be used.
Advance to next parameter.
Carbon Monoxide Signal
A signal from an external carbon monoxide monitor can be used for display of
carbon monoxide content in the Carbon Dioxide/Carbon Monoxide P
–
see Operating Instructions.
Cell T/C? yes & ENT
➞
no & ENT
Flue T/C? yes & ENT
➞
no & ENT
Air T/C? yes & ENT
➞
no & ENT
CO Signal yes
➞
➞
no
Store.
Air Thermocouple
Select whether or not an air thermocouple is to be used:
If air thermocouple is used.
or
If air thermocouple is never to be used.
Advance to next parameter.
Store.
Enter
Enter
Cell T/C? xxxxx & ENT
From
o r
➞
yes
▲
▼
➞
no
CO Signal xxxxx & ENT
To
o r
➞
yes
▲
▼
➞
no
o r
➞
yes
▲
▼
➞
no
Advanced
Analysers
Fuel
? xxxxxxxx xx
From
Continued from previous page.
Continued opposite.
➞
➞
➞