Fig. 4.3C Carbon Monoxide Input, Electrical Calibration
SECURE PARAMETE
Min (CO) mA x x x x x x
Minimum Calibration Value
Set 4mA on the milliamp source.
▼
▲
or
Enter
Store.
Advance to next parameter.
Set the value on the upper display to '0.0'.
Max (CO) mA x x x x x x
Maximum Calibration Value
Set 20mA on the milliamp source.
▼
▲
or
Enter
Store.
Set the value on the upper display to '20.0'.