ABB KM26 Скачать руководство пользователя страница 21

21

KM26 | 

MAGNETIC LEVEL GAUGE  |  OI/KM26-EN REV I

•  magnetic particles in the process fluid have become 

attached to the float changing its buoyancy. Install 
magnet traps in the process connection lines, then clean 
and reinstall float. 

•  solidification of process fluids on the surface of the float 

have changed its buoyancy. Consider replacing with a 
Teflon® (registered trademark of DuPont) ‘S’ coated float, 
heat tracing the chamber to decrease solidification, or 
reduce the amount  of solidifying particles in the 
chamber. 

•  solidification of process fluid on the interior of the 

chamber has decreased the clearance for the float. 
Consider replacing with a coated chamber, heat tracing 
the chamber to decrease solidification, or reduce the 
amount of solidifying particles in the chamber.  

Switch does not work

•  switch installed upside down. Remove and install 

correctly. 

•  float does not travel past the switch during operation. 

Float may encounter float stop prior to activating switch. 
Switch point should be a minimum of 1 in. inside the 
upper and lower stop points for the float. 

•  contacts damaged due to excessive load, inductive load, 

or dead short in the circuit. Replace the switch. 

•  magnet has been demagnetized by proximity to 

magnetic source or ferrous materials. Replace the switch 
and remove the interference.  (continued on next page)

•  distance between the switch and float is too large. Strap 

may be loose, insulation may be too thick, attachment to 
a switch mount rod may have moved or the switch has 
been moved away from the location of a guided float.

•  float has become demagnetized and indicator also 

decouples readily. Have the float re-magnetized at the 
factory  and remove the source of demagnetizing.

 

5     Service and maintenance

Pre-startup

•  If equipment is used in a manner not specified by ABB, 

protection provided by equipment may be impaired. 

•  Manually move the float from 0% to 100% and back to 0% 

prior to startup/check out in order to reinitialize any 
switch accessories (only required if magnetically 
actuated switches are supplied). Switches may 
inadvertently change state during any rough handling 
during transport. 

•  Remove float prior to pressurizing tank / vessel.  
•  Verify the MLGs center-to-center dimension equals that 

of the tank / vessel.

4     Troubleshooting

Indicator decoupling

Causes:

•  float is upside down. Remove, check field strength of 

magnets, and install correctly. Proximity to opposing field 
may weaken magnetic field. 

•  scale assembly is not flat against the chamber due to 

missing straps. Magnetic field strength drops 
exponentially with distance. Add gear clamps to eliminate 
channel separation from chamber. Add  stainless steel 
retaining wires to eliminate indicator tube separation 
from channel. 

•  float stop springs have been bent or broken. Adjust or 

replace springs as needed to prevent float travel outside 
the range of the indicator tube. 

•  scale has been moved allowing float travel outside of 

range or causing too much separation from the float. 
Reposition the scale. 

•  float or indicator de-magnetizing by proximity to other 

magnetic material, high temperature, or repulsive fields. 
Consult factory for re-magnetization of float or replace 
the float and / or indicator and remove the source of 
demagnetization. Sources include floats and switches 
installed upside down, close ferrous  materials, nearby 
magnetic fields, magnetic particles from process piping, 
etc.  

•  indicator tube is no longer sealed and contains moisture 

or dirt. This increases friction inside the tube. Replace the 
tube. 

•  chamber is not vertically level causing increased friction 

between the shuttle and glass and increased distance 
between the float and the scale assembly. Adjust the 
position of the chamber. 

•  indicator tube incorrectly installed. See directions for 

installation. 

•  magnetic particles from the process fluid stuck to float. 

This distorts the magnetic field and changes the float 
buoyancy. Remove and clean float and install magnetic 
traps in the process connections. 

If for some reason, magnetic coupling is lost, it can be 
restored by following three simple steps.

Using a permanent magnet, locate the float inside the 
chamber.

Take the permanent magnet and raise the follower to the 
same level as the float.

Remove the magnet to the side as quickly as possible to 
set the follower in  a spinning motion. 

 

Float sinks or sticks

•  process specific gravity is lower than was specified at the 

time of order. Identify true minimum specific gravity 
requirements considering temperature and pressure 
variations and order new float providing minimum and 
operating specific gravity.  

Float damage may occur if not removed prior to any 
pressure testing. 

CAUTION

Содержание KM26

Страница 1: ...free solution for total and or interface level detection in toxic corrosive high pressure and high temperature processes ABB offers the standard KM26 magnetic level gauge with a chamber of virtually...

Страница 2: ...ts 1 Introduction 2 2 Safety 4 3 Mounting 9 4 Troubleshooting 21 5 Service and maintenance 21 6 MLG parts 28 7 MLG parts ordering 30 8 Customer specific product information 30 9 Certificate EU declara...

Страница 3: ...n the float chamber is a float which is designed to float with approximately 70 to 80 of its mass submerged in the process fluid The float magnet assembly is located such that the magnetic actuation p...

Страница 4: ...installation electrical connection commissioning and maintenance of the KM26 series MLG Qualified personnel are those individuals who have experience in the installation electrical connection commissi...

Страница 5: ...l packaging or other suitably secure shipping method The sender should contact the factory for return authorization number and fill out return form provided at the end of the manual and include it wit...

Страница 6: ...the equipment include but are not limited to limited mechanical agitation shall be used use of stilling wells to mitigate effect of agitation limit rate of change of level to values such that frictio...

Страница 7: ...ttom or lower chamber branch connection locations For the KM26S it is also acceptable to connect a wire ring terminal to a flange stud locked in with another nut The grounding connection must be metal...

Страница 8: ...erature classifications of the ATEX approved KM26 is limited to process temperatures up to 50C below the maximum surface temperature allowed for each of the temperature classes as per the below table...

Страница 9: ...hin 24 hours Do not discard the shipping container until all parts components are verified and checked 3 Mounting General Read the following installation instructions carefully before proceeding Failu...

Страница 10: ...e flange at the bottom and a cap at the top A float assembly hangs below the chamber into the liquid via a guide rod The float assembly consists of a magnet at the top of a guide rod and a float at th...

Страница 11: ...racted to the float not repelled when inserted correctly The KM26T float is wrapped separately in bubble wrap The magnet assembly at the end of the float rod is inserted into the top of the guide cham...

Страница 12: ...delivered correctly set up for measuring the product specified in the order The optional magnetic bar graph indicator is available in yellow black or red white for use in locations where temperature i...

Страница 13: ...ition on the chamber figure 14 Limit switches To signal specific liquid levels the KM26 can be equipped with several different types of ABB limit switches Magnetically actuated limit switches are the...

Страница 14: ...lements overvoltage protection devices lightning protection and or voltage separation capacity must be provided at the plant Check that the existing operating voltage corresponds to the voltage indica...

Страница 15: ...tions gussets support brackets etc may interfere with installation at desired locations Cam actuated switches must be mounted in the correct orientation Magnetic level gague switches Model Approvals E...

Страница 16: ...ring length of the gauge align the zero mark with the zero of the scale if the transmitter measuring length matches the center to center of the gauge align the zero mark with the center of the bottom...

Страница 17: ...tter reads higher than zero raise the transmitter If the transmitter reads lower than zero lower the transmitter 16 Attach proper field wiring to the transmitter according to the instruction manual in...

Страница 18: ...sary to retain the tube in the channel 3 Cryogenic indicator tubes are sealed into the scale assembly and should be replaced as a unit These scale assemblies are strapped in place using gear clamps Ve...

Страница 19: ...the scale is located opposite to the process connections and the insulation will be thinned down to in on that side If the scale has been moved from the position that it is shipped in then the thinne...

Страница 20: ...to identify the true liquid level because of the very abnormal performance Similarly if a large number of small bubbles form underneath and around the float it will decrease the buoyant force that is...

Страница 21: ...f the tank vessel 4 Troubleshooting Indicator decoupling Causes float is upside down Remove check field strength of magnets and install correctly Proximity to opposing field may weaken magnetic field...

Страница 22: ...gh the chamber top end 2 After cleaning the MLG chamber replace the float and drain flange A new flange gasket may be required 3 Note Use gaskets compatible with process fluid 4 Verify that the stainl...

Страница 23: ...Align top tube holder with scale channel and slide down until top surface of the holder is flush with top of scale channel The screw hole should align with the hole in the holder 7 Use a 6 32 machine...

Страница 24: ...24 KM26 MAGNETIC LEVEL GAUGE OI KM26 EN REV I 5 Service and maintenance Figure 27 KM26 shuttle replacement glass installation procedure KM26 shuttle replacement glass installation procedure...

Страница 25: ...older cavity 4 Carefully align replacement glass with shuttle indicator into scale channel and lower into bottom tube holder Ensure glass tube rest flat on rubber bumper and is firmly seated Nipple on...

Страница 26: ...26 KM26 MAGNETIC LEVEL GAUGE OI KM26 EN REV I 5 Service and maintenance KM26 bargraph replacement glass installation procedure Figure 29 KM26 bargraph replacement glass installation procedure...

Страница 27: ...into bottom tube holder Flat faces on glass tube should register with corresponding flat faces in cavity of bottom tube holder Ensure glass tube rest flat on rubber bumper and it firmly seated Nipple...

Страница 28: ...cale assembly includes 1 ABC Ser 1 1A Tube with follower Ser 1A 1B Tube holder Ser 1B 1C Scale Ser 1C 2 KM26S chamber Ser 2KS 3 Vent plug Ser 3KS 4 Gasket Ser 4KS 5 Float chamber closure Ser 5KS Flang...

Страница 29: ...ption Part 1 Scale assembly includes 1 ABC Ser 1 1A Tube with follower Ser 1A 1B Tube holder Ser 1B 1C Scale Ser 1C 2 KM26T chamber Ser 2KT 3 Float stop tube Ser 3KT 4 Float stop plate Ser 4KT 5 Snap...

Страница 30: ...irmed by applications engineering 7 MLG parts ordering 8 Customer specific product information Use this area to record pertinent information about your purchased unit Serial number ___________________...

Страница 31: ...defective the original purchaser shall notify ABB Inc K TEK products and request a returned material authorization before returning the material to ABB Inc K TEK products with transportation prepaid...

Страница 32: ...ata Sheets MSDS and decontamination tags by affixing these documents in close proximity to the shipment label for identification purposes January 18 2006 Return Authorization Form Customer Date Contac...

Страница 33: ...33 KM26 MAGNETIC LEVEL GAUGE OI KM26 EN REV I...

Страница 34: ...34 KM26 MAGNETIC LEVEL GAUGE OI KM26 EN REV I...

Страница 35: ...35 KM26 MAGNETIC LEVEL GAUGE OI KM26 EN REV I...

Страница 36: ...and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prior written consent of ABB 2020 ABB ABB Inc...

Отзывы: