KM26 | 

MAGNETIC LEVEL GAUGE  |  OI/KM26-EN REV I

The installation of insulation on KM26 units is dependent 
upon the process temperatures and type of indicator used.   
All temperatures referenced in insulation instructions refer to 
process temperatures.  For design temperature under 200 °F, 
no insulation is required.  If ordered, the insulation will be a 
tied-on blanket (standard insulation attached to the scale 
assembly).  For 300 °F to 450 °F, an insulation pad is required 
for magnetic bargraph (MBG) indication.  If chamber 
insulation is ordered in lieu of insulation pad, then the 
insulation will be a pipe wrap blanket.  For 450 °F to 500 °F, 
pipe wrap blankets are mandatory for MBG indication.  Before 
installing any blanket or pad, note the location of all chamber 
accessories to ensure a return to the proper working location.  
Some units are built with limited accessory locations and the 
accessories may not function if moved to new location.  

 . . . 3     Mounting

Figure 22 - side and front view of pipe wrap

Edges of the scale assembly can be sharp.  Gloves and 
proper care are required while working.

CAUTION

Insulation pads

To install insulation pads, simply remove the accessory from 
the unit.  Set the accessory on the pad.  Using a razor, cut a 
hole through the pad wherever a bracket or clamp exists.  
Push the brackets or clamps through the hole and re-attach 
the accessory to the previous location.  Verify proper 
operation and location.

The insulation thickness of a blanket is determined by the 
process temperature.  In all locations where scale indications, 
switches, and transmitters are located, the insulation 
thickness is thinned to ½ in. to facilitate magnetic coupling.  
For all other locations the insulation thickness it governed by 
the following:

•  0 to 250 °F (0 to 121 °C) ½ in. thickness ceramic fiber filler 

with inner and outer silicone impregnated cloth covering

•  251 to 500 °F (122 to 260 °C) 1 in. thickness ceramic fiber 

filler with inner and outer silicone impregnated cloth 
covering

•  501 to 850 °F (261 to 454 °C) 1 in. thickness ceramic fiber 

filler with outer silicone impregnated cloth covering and 
4-ply high temp liner including TempKoat™ insulation

•  851 to 1000 °F (455 to 538 °C) 2 in. thickness ceramic fiber 

filler with outer silicone impregnated cloth covering and 
4-ply high temp liner including TempKoat™ insulation

For insulated units, it is critical to know the location of all 
components attached to the unit when ordering the 
insulation.  Typically, the scale is located opposite to the 
process connections and the insulation will be thinned down 
to ½ in. on that side. If the scale has been moved from the 
position that it is shipped in, then the thinned insulation will 
not be in the correct location.  When ordering, ABB refers to 
location by a clock system with the process connections at 12 
o’clock when looking down on the unit from above.  The scale 

is typically at 6 o’clock.  Openings for the connections are 
made on the basis of the orientation specified by the 
customer.  Thinning of insulation is needed for switches and 
transmitters as well and their positions must also be known.

Pipe wrap blanket

For pipe wrap blankets, installation requires the removal of 
scale assembly and transmitter (if installed). If a transmitter 
is installed, note its location and undo the front of the 
brackets (not the gear clamps) and remove the transmitter. 
Switches on switch mount rods will have to be rotated away 
from the chamber.  Wrap the insulation blanket around the 
unit and fasten it with the Velcro straps.  If a transmitter was 
installed, feel for the brackets under the insulation and mark 
their location with a marker.  Sew a 1 in. square through the 
blanket around the marks.  Using a razor, cut an X from corner 
to corner of the square making sure not to break the new 
seam.  Push the brackets through the new holes.  Set the 
scale assembly against the insulation at proper mounting 
location.  Draw a mark on each side of each gear clamp next 
to the scale channel.  This location on the insulation should 
have sewn seams down the length of the blanket to prevent 
fraying.  It is recommend that the marks for each clamp be 
sewn around to reduce fraying.  Using a razor, cut through the 
insulation to allow the gear clamp to pass through the hole.  
Once a hole is made for each clamp, remove the blanket from 
the unit.   Feed the gear clamps of the scale assembly through 
the holes so that each end of the clamp goes through.  Set 
the scale assembly and insulation blanket against the unit 
and use the gear clamps to fasten the parts to the unit.  Make 
sure that the screw for the clamp does not interfere with the 
transmitter or switch locations.  If applicable, feed the 
brackets for the transmitter  through their holes in the 
blanket.  Attach the velcro closures on the back of the blanket.  
Rotate any switches back into position.  Re-attach the 
transmitter at the proper location.  Verify the operation of all 
devices.  An inoperative switch may need to be pressed 
tighter to the unit.  Transmitters should be checked for the 
correct zero point and slid up or down to set to proper level.

Insulation instructions

Содержание KM26

Страница 1: ...free solution for total and or interface level detection in toxic corrosive high pressure and high temperature processes ABB offers the standard KM26 magnetic level gauge with a chamber of virtually...

Страница 2: ...ts 1 Introduction 2 2 Safety 4 3 Mounting 9 4 Troubleshooting 21 5 Service and maintenance 21 6 MLG parts 28 7 MLG parts ordering 30 8 Customer specific product information 30 9 Certificate EU declara...

Страница 3: ...n the float chamber is a float which is designed to float with approximately 70 to 80 of its mass submerged in the process fluid The float magnet assembly is located such that the magnetic actuation p...

Страница 4: ...installation electrical connection commissioning and maintenance of the KM26 series MLG Qualified personnel are those individuals who have experience in the installation electrical connection commissi...

Страница 5: ...l packaging or other suitably secure shipping method The sender should contact the factory for return authorization number and fill out return form provided at the end of the manual and include it wit...

Страница 6: ...the equipment include but are not limited to limited mechanical agitation shall be used use of stilling wells to mitigate effect of agitation limit rate of change of level to values such that frictio...

Страница 7: ...ttom or lower chamber branch connection locations For the KM26S it is also acceptable to connect a wire ring terminal to a flange stud locked in with another nut The grounding connection must be metal...

Страница 8: ...erature classifications of the ATEX approved KM26 is limited to process temperatures up to 50C below the maximum surface temperature allowed for each of the temperature classes as per the below table...

Страница 9: ...hin 24 hours Do not discard the shipping container until all parts components are verified and checked 3 Mounting General Read the following installation instructions carefully before proceeding Failu...

Страница 10: ...e flange at the bottom and a cap at the top A float assembly hangs below the chamber into the liquid via a guide rod The float assembly consists of a magnet at the top of a guide rod and a float at th...

Страница 11: ...racted to the float not repelled when inserted correctly The KM26T float is wrapped separately in bubble wrap The magnet assembly at the end of the float rod is inserted into the top of the guide cham...

Страница 12: ...delivered correctly set up for measuring the product specified in the order The optional magnetic bar graph indicator is available in yellow black or red white for use in locations where temperature i...

Страница 13: ...ition on the chamber figure 14 Limit switches To signal specific liquid levels the KM26 can be equipped with several different types of ABB limit switches Magnetically actuated limit switches are the...

Страница 14: ...lements overvoltage protection devices lightning protection and or voltage separation capacity must be provided at the plant Check that the existing operating voltage corresponds to the voltage indica...

Страница 15: ...tions gussets support brackets etc may interfere with installation at desired locations Cam actuated switches must be mounted in the correct orientation Magnetic level gague switches Model Approvals E...

Страница 16: ...ring length of the gauge align the zero mark with the zero of the scale if the transmitter measuring length matches the center to center of the gauge align the zero mark with the center of the bottom...

Страница 17: ...tter reads higher than zero raise the transmitter If the transmitter reads lower than zero lower the transmitter 16 Attach proper field wiring to the transmitter according to the instruction manual in...

Страница 18: ...sary to retain the tube in the channel 3 Cryogenic indicator tubes are sealed into the scale assembly and should be replaced as a unit These scale assemblies are strapped in place using gear clamps Ve...

Страница 19: ...the scale is located opposite to the process connections and the insulation will be thinned down to in on that side If the scale has been moved from the position that it is shipped in then the thinne...

Страница 20: ...to identify the true liquid level because of the very abnormal performance Similarly if a large number of small bubbles form underneath and around the float it will decrease the buoyant force that is...

Страница 21: ...f the tank vessel 4 Troubleshooting Indicator decoupling Causes float is upside down Remove check field strength of magnets and install correctly Proximity to opposing field may weaken magnetic field...

Страница 22: ...gh the chamber top end 2 After cleaning the MLG chamber replace the float and drain flange A new flange gasket may be required 3 Note Use gaskets compatible with process fluid 4 Verify that the stainl...

Страница 23: ...Align top tube holder with scale channel and slide down until top surface of the holder is flush with top of scale channel The screw hole should align with the hole in the holder 7 Use a 6 32 machine...

Страница 24: ...24 KM26 MAGNETIC LEVEL GAUGE OI KM26 EN REV I 5 Service and maintenance Figure 27 KM26 shuttle replacement glass installation procedure KM26 shuttle replacement glass installation procedure...

Страница 25: ...older cavity 4 Carefully align replacement glass with shuttle indicator into scale channel and lower into bottom tube holder Ensure glass tube rest flat on rubber bumper and is firmly seated Nipple on...

Страница 26: ...26 KM26 MAGNETIC LEVEL GAUGE OI KM26 EN REV I 5 Service and maintenance KM26 bargraph replacement glass installation procedure Figure 29 KM26 bargraph replacement glass installation procedure...

Страница 27: ...into bottom tube holder Flat faces on glass tube should register with corresponding flat faces in cavity of bottom tube holder Ensure glass tube rest flat on rubber bumper and it firmly seated Nipple...

Страница 28: ...cale assembly includes 1 ABC Ser 1 1A Tube with follower Ser 1A 1B Tube holder Ser 1B 1C Scale Ser 1C 2 KM26S chamber Ser 2KS 3 Vent plug Ser 3KS 4 Gasket Ser 4KS 5 Float chamber closure Ser 5KS Flang...

Страница 29: ...ption Part 1 Scale assembly includes 1 ABC Ser 1 1A Tube with follower Ser 1A 1B Tube holder Ser 1B 1C Scale Ser 1C 2 KM26T chamber Ser 2KT 3 Float stop tube Ser 3KT 4 Float stop plate Ser 4KT 5 Snap...

Страница 30: ...irmed by applications engineering 7 MLG parts ordering 8 Customer specific product information Use this area to record pertinent information about your purchased unit Serial number ___________________...

Страница 31: ...defective the original purchaser shall notify ABB Inc K TEK products and request a returned material authorization before returning the material to ABB Inc K TEK products with transportation prepaid...

Страница 32: ...ata Sheets MSDS and decontamination tags by affixing these documents in close proximity to the shipment label for identification purposes January 18 2006 Return Authorization Form Customer Date Contac...

Страница 33: ...33 KM26 MAGNETIC LEVEL GAUGE OI KM26 EN REV I...

Страница 34: ...34 KM26 MAGNETIC LEVEL GAUGE OI KM26 EN REV I...

Страница 35: ...35 KM26 MAGNETIC LEVEL GAUGE OI KM26 EN REV I...

Страница 36: ...and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prior written consent of ABB 2020 ABB ABB Inc...

Отзывы: