![ABB IRB 2400/10 Скачать руководство пользователя страница 163](http://html.mh-extra.com/html/abb/irb-2400-10/irb-2400-10_product-manual_5034163.webp)
5 Calibration information
5.2. Calibration methods
161
3HAC022031-001 Revision: C
©
Co
py
rig
h
t 200
4-
200
8 ABB. All righ
ts reser
v
ed.
Calibration methods
Each calibration method is detailed in a separate manual. Below is a brief description of the
methods available.
Calibration Pendulum - standard method
Calibration Pendulum is the standard method for calibration of all ABB robots (except IRB
6400R, IRB 640, IRB 1400H and IRB 4400S) and is also the most accurate method for the
standard type of calibration. It is the recommended method in order to achieve proper
performance.
Two different routines are available for the Calibration Pendulum method:
•
Calibration Pendulum II
•
Reference Calibration.
The calibration equipment for Calibration Pendulum is delivered as a complete toolkit,
including the operating manual for
Calibration Pendulum
, which describes the method and
the different routines further.
Levelmeter Calibration - alternative method
Levelmeter Calibration is referred to as the alternative method for calibration of ABB robots,
because of the less accurate values obtained during calibration. The method uses the same
principles as Calibration Pendulum but does not have as good of mechanical tolerances to the
toolkit parts as the standard method with Calibration Pendulum.
This method may, after calibration, require modifications in the robot program, and is
therefore not recommended.
The calibration equipment for Levelmeter Calibration is ordered as separate parts for each
robot and the operating manual for
Levelmeter Calibration
is enclosed with the Levelmeter
2000.
CalibWare - Absolute Accuracy calibration
In order to achieve a good positioning in the Cartesian coordinate system, Absolute Accuracy
is used as a TCP calibration. The tool CalibWare guides through the calibration process and
calculates new compensation parameters. This is further detailed in the application manual
Absolute Accuracy Calibware 2.0
.
If a service operation is done to a robot with Absolute Accuracy, a new absolute accuracy
calibration is required in order to establish full performance. For most cases after motor and
transmission replacements that do not include taking apart the robot structure, standard
calibration is sufficient. Standard calibration also supports wrist exchange.
References
Article numbers for the calibration tools are listed in section
The calibration equipment for Calibration Pendulum is delivered as a complete toolkit,
including
Operating manual - Calibration Pendulum
, which describes the method and the
different routines further.
Continued
Содержание IRB 2400/10
Страница 1: ...Product manual Articulated robot IRB 2400 L IRB 2400 10 IRB 2400 16 M2000 M2000A M2004 ...
Страница 2: ......
Страница 8: ...Table of Contents 6 3HAC022031 001 Revision C Copyright 2004 2008 ABB All rights reserved ...
Страница 16: ...Product documentation M2004 3HAC022031 001 Revision C 14 Copyright 2004 2008 ABB All rights reserved ...
Страница 191: ......
Страница 192: ......
Страница 193: ......
Страница 194: ......
Страница 195: ......
Страница 198: ......
Страница 199: ......
Страница 202: ......
Страница 203: ......
Страница 205: ......
Страница 210: ...8 Circuit diagram 8 1 Introduction 3HAC022031 001 Revision C 198 Copyright 2004 2008 ABB All rights reserved Continued ...
Страница 211: ...Manipulator Circuit Diagram 3HAC 6670 3 Rev 01 Product Manual IRB 2400 No of Sheets 13 Sheet no 101 LIST OF CONTENTS ...
Страница 214: ...Manipulator Circuit Diagram 3HAC 6670 3 Rev 01 Product Manual IRB 2400 No of Sheets 13 Sheet no 104 MOTOR AXIS 1 3 ...
Страница 226: ......