4 Startup
Operation
4.1
Preliminary checks prior to start-up
The following points must be checked before commissioning:
• The supply power must be switched off.
• The supply power must match information on the name plate.
• The pin assignment must correspond to the connection diagram.
• Sensor and transmitter must be grounded properly.
• The temperature limits must be observed.
• The sensor must be installed at a largely vibration-free location.
• The housing cover must be sealed before switching on the supply power.
4.2 Operation
4.2.1 Menu
navigation
M00024
4
3
2
1
Fig. 35
• The
◄
(1),
►
(4),
▲
(2) and
▼
(3) buttons are available for the menu-controlled
configuration.
• The menu/submenu name is displayed above in the LCD display.
• The number/line of the currently selected menu item is displayed in the upper right of the
LCD display.
• A scroll bar is located on the right edge of the LCD display which shows the relative position
of the currently selected menu item within the menu.
• Both of the
◄
and
►
buttons can have various functions assigned to them. The meaning of
these buttons is displayed below in the LCD display above the respective button. The
following functions are possible.