28
S4
D184B122U02
Submenu/
Parameter
Submenu/
Parameter
Submenu/
Parameter
Selections
Value Range
Comments
Zero
view only
Adjustment value for Zero can be monitored
for the excitation frequency setting (dis-
played when a new flowmeter primary was
selected).
0 %
Span
→
F
view only
Adjustment value for Span Forward can be
monitored for the excitation frequency set-
ting (displayed when a new flowmeter pri-
mary was selected).
91 %
Span
←
R
view only
Adjustment value for Span Reverse can be
monitored for the excitation frequency set-
ting (displayed when a new flowmeter pri-
mary was selected).
-91 %
Submenu
This menu is not displayed for PROFIBUS
PA.
Data Iout
Adjust Iout 4 mA
view only
Display of the converter adjustment value for
a current output at 4.0 mA.
ENTER
4.0 mA
Adjust Iout 20 mA
view only
Display of the converter adjustment value for
a current output at 20.0 mA.
20.0 mA
Submenu
Primary
Primary Type
Display of the connected flowmeter type.
The selection is automatically uploaded
from the FRAM in the converter. If an older
flowmeter primary model (see list) was se-
lected, then a FRAM with Cs = 100%, Cz =
0% must be used. The Models 10D1422
(10DI1425, 10DS3111A-C
≥
DN 500) can
only be connected to a special converter
connection board designed for these
meters. The switch S903 on the connection
board must be closed. The other flowmeter
models are operated with their assigned
FRAM. Therefore, when older model flow-
meter primaries are connected to the con-
verter the menu "Primary Setup" is
displayed during the initial start-up and in
the parameter “Primary” in the Submenu
“Systemdata” additional parameter settings
can be entered. See also Chapter 10 in the
Operating Instruction.
ENTER
DS2_, DS4_
Mains Frequency
view only
Display of the mains frequency at which the
flowmeter primary was calibrated. Displayed
only for “Older Model” 50/60 Hz. Not dis-
played for the newer flowmeter primaries.
50 Hz