5.3  Disconnecting the sensor electrical power and 

signal cables at the process

Referring to Fig. 5.3:
1.

Use a medium flat-bladed screwdriver to unscrew 4 
(captive) cover screws 

A

 and remove cover 

B

 from 

sensor assembly 

C

.

2.

Disconnect mains cable 

D

 from terminal block 

E

3.

Disconnect signal cable 

F

 from terminal block 

G

.

4.

Disconnect thermocouple cables 

H

 from terminal block 

I

.

5.

If optional blowback is fitted, disconnect cable 

J

 from 

terminal block 

K

.

5.4  Removing the sensor assembly from the process

Referring to Fig. 5.4:
1.

Use a 10 in. adjustable spanner (wrench) to remove 4 nuts, 
washers and lockwashers 

A

 securing sensor assembly 

B

 to mounting flange 

C

. Set items aside for re-use.

2.

Carefully remove the sensor assembly and the attached 
probe (including filter assembly) from the process.

3.

Temporarily cover process opening until the sensor 
assembly is ready to be re-installed.

DANGER – Serious damage to health / risk to life

Allow sufficient time for the sensor assembly to cool 
before performing this procedure.

DANGER – Serious damage to health / risk to life

The transmitter must be isolated from mains power 
supplies before performing this procedure.

Fig.  5.3  Disconnecting sensor cables

Li

n

e

 i

n

N

e

u

tr

a

l in

Ea

rt

h

D–

 (

g

re

e

n

)

S

C

N

 (

s

cr

ee

n

)

D+

 (

w

h

it

e

)

V+

 (

re

d

)

CO

M (

b

la

c

k

Bl

o

w

b

a

c

k

Bl

o

w

b

a

c

k

In

le

t T

/C

 +

In

le

t T

/C

 –

Sc

re

e

n

Ou

tl

e

t T

/C

 +

Ou

tl

e

t T

/C

 –

Ea

rt

h

DANGER – Serious damage to health / risk to life

Allow sufficient time for the sensor assembly to cool 
before performing this procedure.

Fig.  5.4  Removing the sensor assembly from the process

Содержание Endura AZ40

Страница 1: ...nd 8 4 pages 4 4 6 and 12 Replacing sensor assembly flow disk Sections 4 5 6 and 8 5 pages 4 4 6 and 14 Kit contains Orifices gaskets seals This publication Tools required Transmitter door key supplie...

Страница 2: ...contain fibrous material that can be a health hazard if airborne The material predominantly aluminosilicate refractory fibres CAS 142844 00 6 Refractory ceramic fibres RCF are classified as Category 1...

Страница 3: ...sensor probe assembly must be mounted in accordance with the information supplied in Operating instruction OI AZ40 EN Suitable lifting equipment must be available when installing removing the sensor...

Страница 4: ...the sensor air and test gas supplies at the process Referring to Fig 5 2 1 Disconnect air line B and test gas line C at sensor D DANGER Serious damage to health risk to life The transmitter must be i...

Страница 5: ...embly B to mounting flange C Set items aside for re use 2 Carefully remove the sensor assembly and the attached probe including filter assembly from the process 3 Temporarily cover process opening unt...

Страница 6: ...s kit component locations Fig 6 1 Orifice and seals kit component locations on sensor assembly Key A CO orifice see Section 7 page 7 B O2 orifice see Section 7 page 7 C CO heater block gaskets see Sec...

Страница 7: ...a 1 4 in allen key wrench to secure CO orifice pipe plug A in place in the CO orifice chamber 9 Use a 5 32 in allen key wrench to secure O2 orifice D 1 notch in head in place in the O2 orifice chambe...

Страница 8: ...flat bladed screwdriver to unscrew 4 captive cover retaining screws A and remove cover B Retain cover for re use 2 Use a medium flat bladed screwdriver to unscrew 2 captive cover retaining screws C an...

Страница 9: ...crewdriver to depress the spring connector in each terminal and withdraw the wires from the terminal housing 3 Unscrew O2 sensor A from manifold block B using a 7 8 in spanner wrench and withdraw the...

Страница 10: ...th attached washer K 14 Re fit COe block insulation A in the reverse order of removal 15 Apply a light coating of an anti seize compound suitable for temperatures up to 200 C 392 F to sensor clamp pla...

Страница 11: ...Endura AZ40 Sensor assembly Orifice and seals kit INS ANAINST 007 EN Rev B 11 Fig 8 4 Replacing sensor assembly gaskets...

Страница 12: ...ted correctly 7 Fit the nozzle B into the aspirator A ensuring it is seated correctly 8 Apply a light coating of an anti seize compound suitable for temperatures up to 200 C 392 F to the threaded port...

Страница 13: ...Referring to Fig 8 6 If a recommended 2 stage filtration system A is fitted in the instrument air line replace the filters in accordance with manufacturers instructions Fig 8 5 Replacing Cleaning sen...

Страница 14: ...rner of the flow disk assembly J with the upper right hand corner Ensure screw holes are aligned correctly 11 Apply a light coating of anti seize compound suitable for temperatures up to 200 C 392 F t...

Страница 15: ...Endura AZ40 Sensor assembly Orifice and seals kit INS ANAINST 007 EN Rev B 15 Fig 8 7 Replacing sensor assembly flow disk...

Страница 16: ...ts in this document and in the subject matter and illustrations contained therein Any reproduction disclosure to third parties or utilization of its contents in whole or in parts is forbidden without...

Отзывы: