9.2.4 Check Function Messages
Message range: C020.041 to M002.0XX
Status
Icon
Diagnostic Message
Displayed
Possible Cause(s)
Corrective Measure(s)
––C020.041––
SIMULATION
MODE
Signals and/or
diagnostics
being simulated
Transmitter is in simulation mode.
Exit simulation mode before using in
the field.
––C019.044––
FACTORY TEST
MODE
Set to run mode
before using in
the field
Transmitter in special factory test
mode.
Set to normal run mode before using
in the field.
––C018.043––
DEMONSTRATION
MODE
Set to run mode
before using
in the field
Transmitter in demonstration
mode.
Set to normal run mode before using
in the field.
––C016.045––
CURRENT OUTPUT 1
IN TEST MODE
Output under
manual control.
No action req'd
Current Output 1 under manual
control.
No action required.
––C014.046––
CURRENT OUTPUT 2
IN TEST MODE
Output under
manual control.
No action req'd
Current Output 2 under manual
control.
No action required.
Table 9.5 Diagnostics – Check Function Messages