Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
6 Programming
COI/AZ30E–EN Rev. D
33
Signals View
The
Signals View
displays active signals and their values as a list.
Values for signals that have not been configured are not
displayed.
Use the
and
keys to scroll through active signals.
Signals Units
Mnemonic Displayed
Description
Typical
Values
O2
%O2
% Oxygen
0.01 to 100.00 % O
2
TMP
°C or °F
Cell Temperature
>700 °C (1292 °F)
After stabilization phase
dT
°C / Min or °F / Min
Rate of Rise of
0 to 60 °C / min
Cell Temperature
(0 to 108 °F / min)
mV
mV
Cell Millivolts
–50 to 190 mV
O/P
%
Heater Output
0 to 100 %
CJ
°C or °F
Cold Junction Temp
–20 to 70 °C (–4 to 158 °F)
Off
mV
Calibration Offset
–20 mV to 20 mV
Fct
Calibration Factor
0.900 to 1.100
zOf
mV
Cell Zero Offset
<20 mV
Dev
%O2
O
2
Deviation
<0.5 % O
2
K
Cell Impedance
0.1 to 10.0 k
d
K
/ Year
Cell Impedance Drift
<1 k
RT
s
Test Gas Response Time
8 to 30 s
RR
%O2 / s
Recovery Rate
0.05 to 0.30 % O
2
Vac
V
Mains Voltage
90 to 265 V
Frq
Hz
Mains Frequency
50 to 60 Hz
Int
°C or °F
Internal Temperature
–20 to 70 °C (–4 to 158 °F)
Display Overview
Section and Function
Table 6.3 Operator Menu Screens (Continued)
Signals View
Exit
O2 xx.x
%O2
TMP
xxx °C
dT
x.x deg C/m
Back
O2 xx.x
%O2