Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
6 Programming
COI/AZ30E–EN Rev. D
59
…Input/Output
Relay 1 (2)
Polarity
Selects whether the relay contacts are closed or
open if any of the
Relay Assignment
options are
active (
Assigned
).
Active Open
Active Closed
Active Closed
Digital I/O 1 (2)
Digital I/O (1 and 2) parameters are available only if an
option board is fitted.
Mode
Selects if the Digital I/O functions as an input or an
output.
Input
Input
Output
Output Assignment
If the
Digital I/O/Mode
is
Output
, selects which alarm
and diagnostic states are required to energize the
output.
Each
Output Assignment
parameter can be
Assigned
or
Not Assigned
.
Note
.
Assigned
parameters are OR'd together.
Not
Assigned
(all
parameters)
Alarm 1 (2, 3, 4)
See Section 6.4.6, page 57 for alarm details.
Cal. In Progress
Output is triggered when a calibration is in progress.
Cal. Failed
Output is triggered if a calibration fails.
Gas 1 Not Present
Output is triggered if Test Gas 1 is not detected.
Gas 2 Not Present
Output is triggered if Test Gas 2 is not detected.
Parameter
Comment / [Range]
Default
Table 6.12 Input/Output Menu (Continued)