Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
6 Programming
COI/AZ30E–EN Rev. D
45
…Calibrate
…Scheduled Cal.
If AutoCal is fitted, calibrations can be set to run
automatically at scheduled time intervals.
Note
. This option is not available if the
AutoCal
Hardware / Hardware Type
is set to
None
.
Time of Next Cal.
Selects the time and date of the next and subsequent
scheduled calibrations. No scheduled calibrations are
performed until the date / time set is reached. The date
of the next scheduled calibration is updated
automatically according to the frequency set.
For example, if
Frequency
is 5 days and
Time of Next
Cal.
is 12:00:00 2009-01-05, it is updated automatically
to 12:00:00 2009-01-10.
Note
. If the scheduled calibration cannot be run, or is
unsuccessful, the next scheduled calibration date is
updated according to the frequency set and an
Out of
Specification
/
Missed Scheduled AutoCal
diagnostic
message is created – see Section 9.2.2, page 87.
N / A
Sequence
Selects the type of calibration / check to be performed.
Calibration
Performs a calibration that automatically updates the
calibration offset and factor if successful.
Calibration
Accuracy check
Performs an accuracy check – this follows the normal
calibration sequence, but does not update the
calibration offset and factor in use by the transmitter.
Cal. On Error limit
Performs an accuracy check but if the error is greater
than the set
Cal. Error Limit
(see below), the calibration
offset and factor are updated automatically.
Warn On Error Limit
Performs an accuracy check, but if the error is greater
than the
Cal. Error Limit
value, a calibration required
warning is generated.
Cal. Error Limit
Sets the limit for use with
Warn On Error Limit
field.
[0.01 to 10.00 % O
2
*]
1.00 % O
2
*The Endura AZ30 oxygen systems must be used only for measuring non-oxygen enriched gasses (not more than 21 %
oxygen [air]) at pressure no greater than 1.1 bar absolute (44 in. WG). Operating the system in oxygen-enriched
atmospheres will invalidate / compromise certification.
Parameter
Comment / [Range]
Default
Table 6.7 Calibrate Menu (Continued)