Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
6 Programming
COI/AZ30E–EN Rev. D
61
…Input/Output
…Digital I/O 1 (2)
Type
Selects the type of automatic calibrations started by
a digital input – see Section 7.1, page 69.
One point
Two point
Two point
Sequence
Selects the type of calibration / check performed
when activated by a digital input.
Calibration
Calibration
Performs a calibration that updates the calibration
offset and factor automatically if successful.
Accuracy Check
Performs an accuracy check (the normal calibration
sequence) but does not update the calibration offset
and factor in use by the transmitter.
Cal. On Error Limit
Performs an accuracy check but, if the error is
greater than the value set in the
Cal. Error Limit
field
(see page 45), the calibration offset and factor are
updated automatically.
Warn On Error Limit
Performs an accuracy check, but if the error is
greater than the value set in the
Cal. Error Limit
field
(see page 45) a calibration warning is generated.
Parameter
Comment / [Range]
Default
Table 6.12 Input/Output Menu (Continued)