Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
6 Programming
44
COI/AZ30E–EN Rev. D
…Calibrate
…Scheduled Cal.
If AutoCal is fitted, calibrations can be set to run
automatically at scheduled time intervals.
Note
. This option is not available if the
AutoCal
Hardware
/
Hardware Type
is set to
None
.
Daily Interval
Sets the required daily interval for scheduled automatic
calibrations of the probe.
Enabled when
Frequency
is set to
Daily
Daily
Daily
2, 3, 4, 5, 6, 7 Days
Weekly Interval
Sets the required weekly interval for scheduled
automatic calibrations of the probe.
Enabled when
Frequency
is set to
Weekly
Weekly
Weekly
2, 3, 4, 6, 13,
26, 52 Weeks
Monthly Interval
Sets the required monthly interval for scheduled
automatic calibrations of the probe.
Enabled when
Frequency
is set to
Monthly
Monthly
Monthly
2, 3, 4, 6, 12 Months
Parameter
Comment / [Range]
Default
Table 6.7 Calibrate Menu (Continued)