Single and Dual Input Analyzers for Low Level Conductivity
AX410, AX411, AX413, AX416, AX418, AX450, AX455 & AX456
6 INSTALLATION
50
IM/AX4CO Issue 11
6 INSTALLATION
6.1 Siting Requirements
Note.
Mount in a location free from excessive vibration.
Mount away from harmful vapours and/or dripping
fluids.
Where possible, mount the analyzer at eye level to
allow an unrestricted view of the front panel displays
and controls.
Fig. 6.1 Siting Requirements
C – Within Environmental Limits
B – Within Temperature Limits
A – Maximum Distance Between Analyzer and Sensor
Conductivity
Sensor
Maximum Distance
50m (137.5 ft) –
cell constant <0.1
100m (275 ft) –
cell constant
≥
0.1
65
°
C
(149
°
F)
Max.
–20
°
C
(–4
°
F)
Min.
IP66
NEMA 4X