Single and Dual Input Analyzers for Low Level Conductivity
AX410, AX411, AX413, AX416, AX418, AX450, AX455 & AX456
5 PROGRAMMING
IM/AX4CO Issue 11
23
Temperature Compensation Range
Select a temperature compensation range to suit the sample temperature:
Off
Lo TC
Hi TC
-
-
-
Raw conductivity measurement without temperature compensation.
Examples
- Water for injection (WFI) for US Pharmacopoeia (USP) applications.
- Purified water for USP applications.
Temperature compensation for sample temperatures in the range 0 to
100ºC (0 to 212ºF). This setting is suitable for the majority of applications.
Temperature compensation for sample temperatures in the range 0 to
200ºC (0 to 392ºF). This setting is used only for special high temperature
applications.
Continued on next page.
Continued on page 25.
Continued on page 26.
A: Temp.Comp
A: T.Comp Range
-----
Hi TC
Lo TC
Off
A: Temp.Sensor
Hi TC
A: Temp.Comp
Lo TC
Off