5 PROGRAMMING
30
IM/AX4PH
Issue 9
5.3 Configure Sensors
Configure Sensor A
Sensor B configuration (dual input analyzers only) is identical to Sensor A configuration.
Probe Type
Select the required probe type.
ORP
Redox
pH
-
-
-
Millivolt display
Millivolt display
pH display
Differential Input
Yes
-
Select if electrode system is equipped with a solution earth
(ground) rod and sensor diagnostics are required - see section
5.4, page 33.
No
-
Select if electrode system is not equipped with a solution earth
(ground) rod or sensor diagnostics are not required.
Note.
Ensure the electrode system is connected correctly for the type of input selected
(Standard or Differential) - see Fig. 6.9 (wall-/pipe-mount analyzers) or Fig. 6.11
(panel-mount analyzers).
Probe Type
set to
Redox
or
ORP
- continued below.
Probe Type
set to
pH
- continued on next page.
Temperature Sensor
Select the type of temperature sensor used: Pt100, Pt1000, Balco 3K or None.
Enable Calibration
Select
Yes
to enable sensor calibration.
Select
No
to disable sensor calibration and all associated frames.
Return to main menu.
Config. Sensor A
-----
Config. Sensor B
A: Temp. Sensor
A: Probe Type
-----
ORP
Redox
pH
CONFIG. SENSORS
-----
A: Diff. Input
-----
Yes
No
A: Electrode
A: Temp. Sensor
-----
None
3K Bal
Pt1000
Pt100
A: Enable Cals
-----
Yes
No
A: Probe Type
set to
Redox/ORP
CONFIG. SENSORS
CONFIG. ALARMS
Config. Sensor A