Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468
4 SETUP
22
IM/AX4PH
Issue 9
4.1.3 Adjust Offset (Redox/ORP Only)
Calibrate Sensor A
Sensor B (dual input analyzers only) calibration is identical to Sensor A calibration.
Probe Type
for Sensor B (dual input analyzers only) set to
Redox
or
ORP
and
Enable
Cals.
set to
No
(Section 5.3) - return to top of page.
Alter Sec. Code
not set to zero (Section 5.8) - see section 5.1, page 28.
Alter Sec. Code
set to zero (Section 5.8) - see section 5.2, page 29.
Probe Type
set to
Redox
or
ORP
(Section 5.3) - continued below.
Probe Type
set to
pH
(Section 5.3) - see section 4.1.1, page 19.
Adjust Offset (ORP/Redox probes only)
mV
and
Adjust
are shown alternately on the upper display line. Use the
and
keys to adjust the upper display line to the required offset value for the process.
Sensor B (dual input analyzers only) calibration is identical to Sensor A calibration.
Probe Type
for Sensor B (dual input analyzers only) set to
Redox
or
ORP
and
Enable
Cals.
set to
No
(Section 5.3) - return to top of page.
Alter Sec. Code
not set to zero (Section 5.8) - see section 5.1, page 28.
Alter Sec. Code
set to zero (Section 5.8) - see section 5.2, page 29.
Sensor Cal. A
-----
A: Adjust Offset
407
mV
0
mV
A: Adjust Offset
A: Buffer Method
SECURITY CODE
CONFIG. DISPLAY
Sensor Cal. A
SECURITY CODE
CONFIG. DISPLAY
SENSOR CAL.
Sensor Cal. B
SENSOR CAL.
Sensor Cal. B