Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468
5 PROGRAMMING
38
IM/AX4PH
Issue 9
5.6 Configure Outputs
Configure Output 1
Output 2 configuration (and Outputs 3 and 4 if option board fitted
and
analog features
enabled - see section 7.3, page 58) is identical to Output 1 configuration.
Assign
Select the sensor and analog output required:
Sen.A
-
pH/Redox (ORP)/mV for the selected sensor.
Sen.B
Temp.A
-
Temperature for the selected sensor.
Temp.B
A-B
-
Difference between the Sensor A and Sensor B readings.
Note.
Sen.B
,
Temp.B
and
A-B
are applicable only to dual input analyzers.
Range
Select the analog output current range for the selected output.
Span Value
pH
(or
mV
or
Deg.C
or
Deg.F
or
A-B
) and
Adjust
are shown alternately on the upper
display line. Use the
and
keys to adjust the displayed reading to the
required span value.
Note.
The minimum and maximum span values are determined by the
Zero Value
setting (see next page) plus the minimum differential, e.g. to set
Span Value
to 0.00 pH,
first set
Zero Value
to -2.00pH.
pH
Redox/ORP
Temperature
A-B
-
-
-
-
0.00 to 16.00 pH (minimum differential, 2.00pH)
-1100 to 1200 mV (minimum differential, 100mV)
Deg. C 0.0 to 150.0 (minimum differential, 10ºC)
Deg. F 32.0 to 302.0 (minimum differential,18ºF)
0.00 to 14.00 pH (minimum differential, 2.00pH)
Note.
A-B
is applicable only to dual input analyzers.
Continued on next page.
Config. Output 1
-----
AO1: Assign
-----
AO1: Range
-----
CONFIG. OUTPUTS
-----
A-B
Temp.B
Sen.B
Temp.A
Sen.A
4-20mA
0-20mA
0-10mA
AO1: Zero Value
Config. Output 2
AO1: Span Value
7.00
pH
0.00
pH