Single and Dual Input Analyzers for pH/Redox (ORP)
AX416, AX436, AX460, AX466 & AX468
7 CALIBRATION
IM/AX4PH
Issue 9
61
Calibrate Output 1
Note.
When adjusting the 4 and 20mA outputs, the display reading is unimportant and
is used only to indicate that the output is changing when the
and
keys are
pressed.
See below.
Adjust 4mA
Use the
and
keys to set the milliammeter reading to 4mA.
Note.
The analog output range selected in
Configure Outputs
(Section 5.6) does not
affect the reading.
Adjust 20mA
Use the
and
keys to set the milliammeter reading to 20mA.
Note.
The analog output range selected in
Configure Outputs
(Section 5.6) does not
affect the reading.
See below.
Operating Page
Calibrate Output 2
Note.
Output 2 calibration is identical to Output 1 calibration.
Option board fitted
and
analog features enabled - continued on next page.
Option board fitted, additional features disabled - continued on next page.
Option board not fitted - continued on next page.
Option board fitted
and
analog features enabled - continued on next page.
Option board fitted, additional features disabled - continued on next page.
Option board not fitted - continued on next page.
Operating Page
Cal. Output 1
-----
O1: Adjust 4mA
16000
O1: Adjust 20mA
7200
Monitoring pH
Cal. Output 2
Cal. Output 1
Cal. Output 2
Cal. Output 2
-----
O1: Adjust 20mA
7200
Monitoring pH
Cal. Output 3
Cal. Output 2
Cal. Output 3
Option Board
Option Board
Alter Fact. Code
Alter Fact. Code
Redox (ORP)
Dual Redox
pH/Redox (ORP)
Redox (ORP)
Dual Redox
pH/Redox (ORP)