![ABB ACF-NT Скачать руководство пользователя страница 17](http://html.mh-extra.com/html/abb/acf-nt/acf-nt_operators-manual_4305017.webp)
42/23-572 EN Rev. 5
Operator’s Manual ACF-NT, ACF-NT V0309
17
“ASP Module” Screen
Figure 6
“ASP Module” Screen
Indication
The “ASP Module” screen displays the values of the ASP module temperature and
pressure regulators.
Parameter
Default
T-Co.D
Temperature controller sample conditioning block
180 °C
T-Co.E
Temperature controller sample cell
180 °C
Input
Sample cell pressure
850 hPa
Output
Injector pump pressure
800 hPa
Note The
Multi-FID
14
T-Co.D
value actually belongs to the “Multi-FID” screen (see
page 18).