42/23-572 EN Rev. 5
Operator’s Manual ACF-NT, ACF-NT V0309
37
VOC Analyzer Calibration
Zero Calibration
For the zero calibration either nitrogen from a cylinder or the zero gas for the
oxygen analyzer is used. In the latter case a T-piece must be used to connect the
test gas cylinder with the test gas port and the FID zero gas port of the analyzer
system.
Span Calibration
For the span calibration n-Propane in nitrogen from a test gas cylinder is used.
Maintenance Interval
According to the officially established maintenance interval zero and span calibra-
tion have to be performed every 14 days.
Calibration Control
The VOC analyzer calibration is performed as externally controlled automatic
calibration.
The FTIR controller triggers the start of the automatic calibration.
Enable Off
Test gas concentration
Zero
0
Span
Bottle
concentration
Single zero calibration
Never
Single span calibration
Never
Calibration Data
Common zero and span calibration Always