38
Operator’s Manual ACF-NT, ACF-NT V0309
42/23-572 EN Rev. 5
Status Messages
No. Message
Text
F MR
MM Comments
Syst. Failure
X
Measured value is invalid.
Syst. Maint. Req.
X
Measured value is valid.
Syst. Maintenance
X
Discard the measured value as a process measured value.
Module
Failure
Error
on
Module RGM11, MultiFID14, FTIR.
FID Error
FID measured value is invalid.
FID Maint. req.
X
The analyzer is in a state that will soon require user
intervention.
FID Maint. mode
X
The analyzer is being calibrated or serviced.
O2 Error
O2 measured value is invalid.
O2 Maint. req.
X
The analyzer is in a state that will soon require user
intervention.
O2 Maint. mode
X
The analyzer is being calibrated or serviced.
110 Sys.
Boot
1001 Pump
ctrl.
10000 Emerg.
Purge
X
Emergency
purge,
check temp error (FTIR, probe, pipe,
sample gas line).
10001 Message back purge
X
10002 FTIR Failure
X
Contact service.
10003 FTIR Maint. Req.
X
Contact service.
10004 FTIR
Maintenance
X
10005 Temp MGL
X
Temp. error sample gas line.
10005 Temp Pipe
X
Temp. error pipe.
10005 Temp Probe
X
Temp. error sample probe.
10007 Flow spectrom.
X
Flow to FTIR is too low.
10008 Service
X
Key “Service” is activated.
10009 Enablir failure
X
Contact service.
10010 Enablir
Maint.
Req.
X
10011 Enablir
Maintenance
X
10013 Com watchdog
X
Contact service.
10014 Sequence trigger
X
Key “Shut down” is activated.
10015 Zero Y04
X
Zero gas valve is activated.
10016 Span Y05
X
Span gas valve is activated.
10017 Error save data
10018 Shut down
Shut down process is running.
10019 High conc.
X
Concentration too high.
10020 Purge
locked
10021 Start Purge
X
Key “Start purging” is activated.
10022 ZRO2
X
Detector measuring value exceeds the –10000 to +5000
range. Oxygen-free test gas is being used.
F =
Failure
MR = Maintenance Request
MM = Maintenance Mode