Cleaning/Replacing the Sample Gas Filter in the SC Block
When is cleaning /
replacement needed?
Clean or replace the 1 μ stainless steel filter in the sample conditioning block
(SC block) if it is contaminated and the sample gas flow is reduced.
To avoid a prolonged down time of the analyzer system the complete gas filter
ACF-NT (Part No. 0768914) should be changed. The disassembling, cleaning and
assembling of the used filter should be done separately.
CAUTION!
The sample conditioning block is hot (approx. 180 °C).
Step Action
1
Turn off the sample gas supply.
Loosen the three mounting screws
1
(4 mm hex key) and remove the
sample gas filter cover
2
from the sample conditioning block.
Cleaning / Replacing
the Sample Gas Filter
2
1
2
Continued on next page